Defi 5 AED with Lock-out Protection to Prevent Inadvertent Defibrillation, Automated External Defibrillator
$1,570.00$1,950.00 (-19%)
FREE SHIPPING! Delivery in 3 Days
Xpress is a defibrillator Monitor with voice alarm which can meet your clinical needs from Basic Life Support (BLS) to Advanced Cardiac Life Support (ACLS).
Automatic (Atutomated) External Defibrillator features include:
AED Storage Software (Option)
Its used to assist upper supervisory system of Automatic External Defibrillator DEFI and support transmission of storage data from lower control system of DEFI. It is used for event documentation, display of real-time ECG, voice prompts waveform, display of voice prompts and print of selected waveforms.
Physical
Category | Specification |
Dimensions | 303 x 216 x 89 mm |
Weight | 2.0 kg |
Operating Temperature | 0℃ to 40℃ |
Operating Humidity | Relative humidity between 30% and 95% (non-condensing) |
Storage Temperature (without battery) |
-20℃ to 55℃ |
Storage Humidity (without battery) |
Up to 93% (non-condensing) |
Electrical Isolation
Category | Specification |
Power | Unit operates on internal battery only |
External Electrical Connections | No external devices are attached to the unit |
Risk Current Category | Internally powered equipment with defibrillator-proof BF type patient applied part (as per definition of IEC 60601-1 standard) |
Defibrillator
Specification |
Biphasic Truncated exponential |
150, 150, 200J |
10 sec. to 150J 15 sec. to 200 J |
9 sec. |
Less than 30 seconds |
Less than 35 seconds |
Less than 25 seconds |
22 audible prompts |
LED prompts |
Two buttons – On/Off, Shock |
±15% into any impedance from 25 to 175Ω |
20Ω to 200Ω |
Waveform details
The table below provides details of the biphasic truncated exponential waveform delivered by the DEFI5 (set to 200J) when connected to resistive loads of 25 through 175 Ohms. The waveforms are characterized by typical values for peak current (Ip), duration of the first output phase, and duration of the second output phase. Values are within 10%.
Output Energy (J) |
Patient Impedance (Ω) |
Ip1 (Amps) |
Ip2 (Amps) |
Phase1 (ms) |
Phase2 (ms) |
Interval (ms) |
200 | 25 | 53.6 | 36.0 | 5.5 | 3.3 | 0.7 |
50 | 27.8 | 17.5 | 10.9 | 6.6 | 0.7 | |
75 | 18.2 | 12.0 | 16.3 | 8.9 | 0.7 | |
100 | 13.4 | 9.0 | 19.7 | 9.6 | 0.7 | |
125 | 10.5 | 7.2 | 20.5 | 11.2 | 0.7 | |
150 | 8.6 | 6.0 | 21.6 | 10.4 | 0.7 | |
175 | 8.2 | 6.1 | 21.8 | 11.2 | 0.7 |
Standard:
AED Pad    ……………….. 1Pcs
Battery  ………………………1Pcs
AED Case  …………………. 1Pcs
Cleaning Cloth  …………….1Pcs
Options:
AED Pad for Child
AED Cabinet
IrDA USB Adapter for ECG Data transfer
CD Software for ECG Data transfer